Operational Characteristics
SureTemp
®
Model 678/SureTemp
®
Model 679 1-9
6. Instant Anti-Theft
Audio Alarm System
(Model 678 only)
The user can select an anti-theft option that instructs the
instrument to beep continuously (beginning after
approximately 5 seconds) whenever the unit is removed from
the wall holder.
The user must defeat the Instant Anti-Theft Audio Alarm
within 30 seconds after removal from the wall holder by
holding the Mode button and removing the probe from
the probe well, or by returning the unit to the wall holder.
After 30 seconds, the instrument must be returned to the
wall holder to silence and reset the Instant Anti-Theft
Audio Alarm.
Once in the Instant Anti-Theft Audio Alarm state, the
instrument will continue the audio alarm until power is
exhausted or until the unit is returned to the wall holder.
Note:
At power-up, this mode is deactivated.
7. Error Log (Function 5
in the Model 679)
The instrument will save the last 10 error messages that
occur and display those messages in a last in first viewed
sequence. While in error log, the display reads E x.x. To
display the next error, momentarily press the mode button.
Note:
Battery removal will clear the error log of all error
messages.