Operational Characteristics
SureTemp
®
Model 678/SureTemp
®
Model 679 1-7
F/C Conversion
When a final temperature is displayed (in Normal, Recall, or Monitor mode), pressing
and releasing the Mode button will toggle the temperature between °F/°C.
Note:
A recalled temperature will be displayed in whichever scale (°F/°C) is selected
at the time of recall. This can also be changed during display.
If pressing the Mode button does not change the scale of the displayed temperature,
refer to
Mode Button Problems
in the
Troubleshooting
section.
Temperature Recall
Whenever the instrument is in Low Power or Pulse Timer mode, pressing and
releasing the Mode button will cause the most recent predicted temperature to be
displayed for 5 seconds. An “A”, “O”, or “R” will be displayed near the lower right
corner of the LCD, designating Axillary, Oral and Rectal temperatures for the recalled
temperature. Pressing the Pulse Timer button will interrupt the temperature recall
function.
Note:
No Monitor temperatures will be saved for recall. When a temperature is
recalled, the mode in which it was obtained (axillary, oral or rectal) will be
shown independent of the present mode.
If the last temperature can not be recalled, refer to
Temperature Recall Problems
in
the
Troubleshooting
section.
Biotech Mode
To enter this special program mode:
1. Place the thermometer in the wall holder.
2. Press and hold the Mode button, and at the same time remove the probe from the
probe well.
Press the Mode button approximately 2 seconds to move sequentially through the
various program categories. Changes within each category can be made by
momentarily pressing and releasing the Mode button.
Upon entering the Biotech mode, the following features and selections become
available. A number is provided in the flag area to indicate which Biotech category is
displayed.
To exit Biotech mode at any time, insert the probe into the probe well. Replacing the
probe into the probe holder returns the unit to normal operation. The Biotech mode
will automatically time out after 5 minutes of inactivity.