SureTemp Plus Module OEM Implementation Communications Protocol
Welch Allyn OEM Technologies Confidential 53
(9435) Set Default Thermometer Mode
Causes the module to read the default thermometer mode from the host. The default mode
specifies whether the device enters normal predictive mode or monitor mode when the probe
is pulled.
Command Code (hex)
9435
Data
Size (bytes) Type Description
1 int8 0: Normal (Predictive)
1: Monitor
Response Data
Size (bytes) Type Description
0 N/A None
Notes
Use the “Save Calibration Data to EEPROM” command (page 41) to save permanently to
EEPROM.