
Speci
es
Code Definition / Description Origin/Alert
type
Traditional
code map
(ST+ OEM)
Traditio
nal code
map
(ST+
HH)
Details Host
controller
action
Technical
Service.
5 24 The module’s power rail
is too low.
Power/
Equipment alert
27 Internal
malfunction.
Contact Welch
Allyn
Technical
Service.
Indicate
module
error
5 25 The module’s power rail
is too low.
Power/
Equipment alert
28 Internal
malfunction.
Contact Welch
Allyn
Technical
Service.
Indicate
module
error
5 26 The reference voltage
circuit was detected to
be under voltage or
unstable.
Module/ Host
controller alert
61 E4.2 Internal
malfunction.
Contact Welch
Allyn
Technical
Service.
Indicate
module
error
8 1
Measurement is below
allowable temperature
values and beyond the
low ambient and/or
patient limits.
Ambient/ User
Alert
45 E0.2
Verify
conditions at
26.67° C (80°
F) and 10°C
(50° F),
respectively.
If conditions
are valid and
the problem
persists,
replace the
probe. If
problem
persists,
replace
module.
Clear
8 2
Measurement above
allowable temperature
values and beyond the
high ambient and/or
patient limits.
Ambient/ User
Alert
46 E0.2
Patient or
environmental
temperature
conditions
may be too
high. Verify
conditions at
43.33° C
(110° F) and
40° C (104°
Clear
Commenti su questo manuale