Welch Allyn 12500 Manuale di Servizio

Navigare online o scaricare Manuale di Servizio per Alimentatori Welch Allyn 12500. Welch Allyn 12500 Service manual Manuale Utente

  • Scaricare
  • Aggiungi ai miei manuali
  • Stampa

Sommario

Pagina 1 - Traditional Instruments

Traditional InstrumentsService ManualService ManualTraditional InstrumentsPN 117081-1Medical Division4341 State Street Road, P.O. Box 220Skaneateles

Pagina 2

D 11500 and 11511 2.5 Volt Ophthalmoscope Reassembly:D7 Lens Disc Ass'y -- Place lens disc assembly onto hub (number side up) and engagein dete

Pagina 3 - TABLE OF CONTENTS

TOOLS / MATERIALS / SUPPLIESTweezers (Grobet Peer #7 Stainless Steel or Equivalent)Miller Stephenson Aero-Duster #MS-222 (or Equivalent)Soldering Iro

Pagina 4

Section 10EPISCOPEMODEL # PAGE47300 1-4

Pagina 5

EpiscopeClean all returned Episcopes by wiping with a solution of 10%CLOROX / 90% WATER. Usespecial tools, lens cleaner, and lens tissue as listed in

Pagina 6

47300 Disassembly Continued6. Lamp Collar — (Removal will cause damage to lamp collar. Replacewith new collar). Place a thin flat blade screwdriveri

Pagina 7

EPISCOPETOOLS/MATERIALS/SUPPLIESBayonet Pliers ( Welch Allyn T-10802 )Strap Wrench ( Welch Allyn T-10913 )Tweezers (Stainless Steel or Plastic, Must

Pagina 8

Inspections For Repaired Episcopes:1. Attach the Episcope to a Welch Allyn 3.5volt power source.2. Activate the scope and adjust to maximum light inte

Pagina 9

Section 11LARYNGOSCOPE HANDLESMODEL # SUB-SECTION PAGE60200 A 160300 B 1-260305 C 2-360400 D 360700 E 3-460710 E 3-460720 E 3-460713 F 4-560803 G 5608

Pagina 10

Laryngoscope HandlesRefer to diagrams in current Repair Parts Catalog 900299A. 60200 Disassembly:A1. Hook-on Neck — Unscrew bottom cap. Remove batteri

Pagina 11

60300 Disassembly ContinuedB3. Bottom Cap Spring — Unscrew bottom cap. Unscrew spring.** 60300 Reassembly **B4. Bottom Cap Spring — Screw in replacem

Pagina 12

60305 Reassembly ContinuedC6. Hook-on Neck — Assemble hook-on neck sub-assy onto handle tube andtighten. Insert set screw and tighten.C7. Batteries

Pagina 13

E 11600 3.5 Volt Halogen Ophthalmoscope Disassembly:E1 Bumper -- Pull corner, peel back.E2 Back Cover -- Loosen retaining ring.(Doctor-Side)E3 2 Scr

Pagina 14

** 60700, 60710, 60720 Reassembly **E5. Bottom Cap Spring — Screw in replacement spring. Assemble insulator ontop of the spring.E6. Top Cap Sleeve — A

Pagina 15

60713 Disassembly ContinuedF4. Lamp — Remove switch plunger (resembles typical lamp hood)by unscrewing. Unscrew lamp.** 60713 Reassembly **F5. Lamp —

Pagina 16

H. 60804 Disassembly:H1. Batteries — Unscrew bottom cap and remove 2 “AA” batteries andinner tube assembly.H2. Spring — Unscrew spring in bottom cap.H

Pagina 17

60813, 60814 Disassembly ContinuedI4. Lamp — Remove switch plunger (resembles typical lamp hood)by unscrewing. Unscrew lamp.** 60813, 60814 Reassembl

Pagina 18

60815 Disassembly ContinuedJ4. Lamp — Remove switch plunger (resembles typical lamp hood)by unscrewing. Unscrew lamp.** 60815 Reassembly **J5. Lamp —

Pagina 19

Laryngoscope HandlesTOOLS/MATERIALS/SUPPLIESRetaining “O” Ring Insertion Tool (Welch Allyn T-9583)Insulator Assembly Driver (For 60400) (Welch Allyn T

Pagina 20

Section 12ILLUMINATORMODEL # PAGE60500 1-2

Pagina 21

ILLUMINATORSpecial Instructions: Refer to corresponding diagrams in Welch Allyn Repair Parts Catalog.60500 Disassembly:——————————————————————————————

Pagina 22

60500 Reassembly Continued7. Lamp — Clean the lamp and insert into the nose assembly. Assemblenose assembly to handle tube.8. Batteries — Insert 2 “

Pagina 23

Section 13LARYNGOSCOPE BLADESMODEL # PAGE68040 1-268041 1-268042 1-268043 1-268044 1-268045 1-268060 1-268061 1-268062 1-268063 1-268064 1-268065 1-26

Pagina 24

F 11605 3.5 Volt Halogen Ophthalmoscope Disassembly:F1 Bumper -- Pull corner, peel back.F2 Back Cover -- Loosen retaining ring.(Doctor-Side)F3 2 Sc

Pagina 25

Instructions for Cleaning, Disinfecting and Sterilizing Laryngoscope Blades prior to Service WorkReusable Laryngoscopes (Fiber Optic and Standard)634,

Pagina 26

COMPATIBLE NOT COMPATIBLEInspection Instructions: (After Disinfection)•Inspect for any obvious defects or missing component(s).•Temporarily install co

Pagina 27

Disassembly/ReassemblyA Blade Disassembly:1. RemoveLight carrier Remove locking screw assembly by rotating counter clockwise.(use flat-blade screwdri

Pagina 35

G 11610 3.5V Halogen Coaxial Sealed Ophthalmoscope Disassembly:G1 Bumper -- Pull corner, peel back.G2 Back Cover -- Loosen retaining ring. (Doctor-Si

Pagina 42

Section 14BATTERY HANDLESMODEL # SUB-SECTION PAGE70000 A 170100 A 170500 A 170110 B 270150 C 270160 C 270700 D 370720 D 371000 D 371000-C D 371020-C D

Pagina 43

BATTERY HANDLESSpecial Instructions: Refer to corresponding diagrams in Welch Allyn Repair Parts Catalog 900299.Battery Handles- 70000, 70100, 70110,

Pagina 44

B. 70100, 70110 Disassembly:B1. Batteries — Unscrew bottom cap assembly. Remove rechargeablebattery.B2. Pocket Clip — Unscrew lock nut. ( Use spanner

Pagina 45

D. 70700, ( 70720 IEC ) 2.5V Rechargeable Handle Disassembly:71000, 71000-C, 3.5V Rechargeable Handle Disassembly:71020-IEC, 71020-C 3.5V Rechargeable

Pagina 46

H 11620 3.5V Halogen Coaxial Sealed Ophthalmoscope Disassembly:H1 Bumper -- Pull corner, peel back.H2 Back Cover -- Loosen retaining ring. (Doctor-S

Pagina 47

E. 71500 2.5V, 71670 3.5V Rechargeable Handle Disassembly:E1. Battery — Unscrew bottom cap assembly. Remove rechargeablebattery.E2. Top Cap Assembly —

Pagina 48

E5. a)Bottom Cap Insulator __ Screw bottom cap insulator into bottom cap. Insertcontact plunger, contact spring, button contact. Screwhollow screw

Pagina 49

F. 72801 2.5V Pocket Set Handle Disassembly:F1. Battery — Unscrew bottom cap assembly. Remove rechargeablebattery or 2 ‘AA' batteries.** 72801 Re

Pagina 50

Section 15LIGHT HANDLESMODEL # SUB-SECTION PAGE73200 A 173210 B 273305 B 278000 C 2

Pagina 51 - 18200 streak and 18300 spot

LIGHT HANDLESSpecial Instructions: Refer to corresponding diagrams in Welch Allyn Repair Parts Catalog PN 900299.Light Handles - 73200, 73210 With 73

Pagina 52

B. 73210 Light Handle, 73305 Transformer Disassembly:B1. Halogen Lamp — Unplug transformer from power source. Unplug lighthandle cord from transforme

Pagina 53

LIGHT HANDLESTOOLS/MATERIALS/SUPPLIESPin Vise “C” Size ( L.S.Starrett Co.166C ) for Drills/TapsTweezers (Stainless Steel) Offset 4 1/4" Long (Mo

Pagina 54

Section16PEN LIGHTSPOCKET LIGHTSMODEL # SUB-SECTION PAGE76000 A 1-276600 B 2-377700 C 3-477800 D 5-677900 E 6-8

Pagina 55

PEN LIGHTSPOCKET LIGHTSSPECIAL INSTRUCTIONS:Refer to corresponding diagrams in Welch Allyn Repair Parts Catalog PN 900299. Theterms switch assembly,

Pagina 56

76000 Reassembly ContinuedInsert into head: plunger spring, cup retainer with raisedring facing plunger spring just inserted. Assemble longcup retaine

Pagina 57

I 11630 3.5 Volt Halogen Coaxial™ Ophthalmoscope(With Multiplier) Disassembly:I-1 Bumper -- Pull corner, peel back.I-2 Back Cover -- Loosen retaini

Pagina 58

** 76600 Reassembly **B3. Batteries — Insert 2 “AAA” batteries into the switch/tube assembly.(+ End facing lamp end).B4. Lamp — Screw new lamp onto sw

Pagina 59

77700 Reassembly ContinuedInstall replacement detent spring onto contact screwand insert in switch housing (with the open end of thespring located ov

Pagina 60

D. 77800 Pocketlight Disassembly:D1. Batteries — Unscrew focusing adaptor assembly. Remove 2 “AA”alkaline batteries.D2. Lamp — Lift lamp out of foc

Pagina 61

77800 Reassembly ContinuuedAssemble the switch assembly by lowering it downover the retaining nut. Secure by screwing retainingnut up into the switc

Pagina 62

77900 Disassembly ContinuedE4. Plunger Release — Remove: retaining nut, cup contact-longs, contactspring, cup contact-short, sleeve. Clamp cap in s

Pagina 63

77900 Reassembly ContinuedE7. Lamp — Clean the lamp and screw it into the nose. Slide thelamp hood over the lamp. It must fit tightly to preventit

Pagina 64

TOOLS/MATERIALS/SUPPLIESTweezers (Stainless Steel or Plastic, Must Be Non-Magnetic, Grobet Peer # 7 )12"X 12" (Approximately) Foam Pad (Mous

Pagina 65

1B. I. O. Model 12000Section 17Binocular Indirect OphthalmoscopeModel# Model Name Section Page12000 B.I.O. 1 3-774101 Battery Pack 2 874141 B.I.O. Wal

Pagina 66 - TROUBLESHOOTING OTOSCOPES

2Special Instructions for B.I.O. ServiceClean all returned items by wiping with a solution of 10% Clorox/90% water. Do not use alcoholor other chem

Pagina 67

3Disassembly of Model 12000 B.I.O.Refer to Figure 120103 in Repair Parts CatalogAbbreviations used in text and parts list:nnnnnCMCentral MirrornnnnnEM

Pagina 68

I 11630 Ophthalmoscope Reassembly ContinuedI-11 Cover -- Place cover/disc assembly onto patient-side cover/optical(Doctor-Side) tube base assembly. T

Pagina 69

41.3 Illumination System Place BIO in fixture T-12198 to ease further(IS) Sub-Assembly disassembly.Remove the two central screws that were loosenedin

Pagina 70

5Reassembly of Model 12000Refer to Figure 120103 in Repair Parts CatalogImportant: Check alignment prior to reassembly of whole unit.1.8 Illumination

Pagina 71

6Reassembly of Model 12000 cont’d1.9 Optics Carrier Place optics carrier (ears up), over lens of aperture spacerIS lenses and assembly (lower lens.)c

Pagina 72 - 26000 1-3

7End Reassembly1.12VS Housing cont’dnnnnn Note: Lower two screw heads must not interfer with slider motion.Insert the tabs of new CMH button in CM s

Pagina 73

8Model 74201 Battery PackEnd Disassembly of 74201 Battery PackEnd of Reassembly of 74201 Battery PackReassembly2.1 Back Case Unscrew and remove two sc

Pagina 74

9Model 74241 Wall Bracket Disassembly3.1 Back Cover Remove two screws at back of wall bracket.Pull back cover off at the screw end. It will hinge at

Pagina 75

103.6 Back Cover Engage tabs and back cover into slots of case.Close back cover onto case.Operate black actuator to assure that it is operating micro

Pagina 76

11Required Tools for BIOGeneral purpose tools:3/32” hex keyPeer #7 tweezers#2 phillips screwdriver (Cresent #2501)Long nose pliers (Xcelite 91 cg)Welc

Pagina 79 - 28100 B 2-3

J 11650 Ophthalmoscope Reassembly ContinuedJ11 Lens Disc Assy. -- Clean disc. Place disc on hub.(Number side down, engaged with detent)J12 Cover -- P

Pagina 80

1Section 18Welch Allyn B.I.O 2 Model 12500Binocular Indirect OphthalmoscopeIndexModel# Model Name Section Page12500 B.I.O. 2 1 2 - 874201 B.I.

Pagina 81

2Special Instructions for B.I.O. 2 ServiceClean all returned items by wiping with a solution of10% Clorox / 90% water. Do not use alcoholor other

Pagina 82

3Abbreviations used in text:•BISB.I.O. 2 subassembly without headband•CMCentral mirror•EMEye mirror•CMHCentral mirror holder•IMIllumination mirror•VSV

Pagina 83

41.3IS Ass’y Place B.I.O. 2 in nest fixture T-12198.Remove the two central chassis screws loosened in step1.2.Separate the IS assembly from the VS ass

Pagina 84

5Abbreviations used in text:•BISB.I.O. 2 subassembly without headband•CMCentral mirror•EMEye mirror•CMHCentral mirror holder•INIllumination mirror•VSV

Pagina 85

61.8 Illumination System Bracket IS Lenses Cont’dNote: use plastic tweezers to hold lenses and wide band hot mirrorPlace IM yoke and mirror asse

Pagina 86

71.11 Heat Sink Lamp Holder Place contact holder onto lamp socket of the heat sinklamp holder.Lower this assembly onto the optics carrier.Pick up t

Pagina 87

81.14 Lamp Cover Assembly Inspect that the lamp is fully into lamp housing.Slide the lamp cover down into the assembly.Installation of:1.15 Head Band

Pagina 88

9Disassembly of Model 74201 Battery Pack (B.I.O. 2)2.1 Back case Unscrew and remove two screws from back of batterypack .Carefully separate back

Pagina 89

10Disassembly of model 74241 Wall Bracket (B.I.O. 2)3.1 Back cover Remove two screws at back of wall bracket.Pull back cover off at the screw end. It

Pagina 90 - Model #:

K 11660 3.5 Volt Halogen Coaxial™ Autostep Sealed Ophthalmoscope with Multiplier:Disassembly and reassembly for the 11660 are the same as for the

Pagina 91

11Reassembly cont’d3.6 Back cover Engage tabs of back cover into slots of case.Close back cover onto case.Operate black actuator to assure that it is

Pagina 92

12Required Tools for B.I.O.2 ServiceGeneral purpose tools:3/32" hex keyPeer #7 tweezers#2 phillips screwdriver (Crescent #2501)Long nose pliers (

Pagina 95

1CompacSet Components:21110 2.5 Volt Otoscope12810 2.5 Volt Ophthalmoscope74001 CompacSet, Handle OnlyRefer to corresponding diagrams in Welch

Pagina 96

2The CompacSet requires either two AA Alkaline batteries or the Welch Allyn rechargeablebattery PN 72610.Remove battery cover and replace batteries

Pagina 97 - for 49003 & 49022

36. Lens -- Clean lens and slide lens into Otoscope Body Assembly.7. Lamp Collar -- Insert replacement lamp collar into Otoscope IlluminatorAssembly

Pagina 98

4Special Instructions:Clean all returned Ophthalmoscopes by wiping with a solution of 10% Clorox / 90%Water.Do not apply alcohol, chemicals or water t

Pagina 99

5D. 12810 2.5 Volt Ophthalmosccope Disassembly, cont’d5. Ophthalmoscope -- Remove N.D. (Number Dial) filter assembly. (On newerBody Assembly models

Pagina 100

61. Operate all slide switches, rotating knobs and discs to determine whether or not detentsare operating correctly and to determine if the part opera

Pagina 101 - 47300 1-4

M 11720 3.5 Volt Coaxial Sealed Ophthalmoscope Disassembly:M1 Bumper -- Pull corner, peel back.M2 Back Cover -- Loosen retaining ring. (Doctor-side)M

Pagina 102

75. Check windows for cracks, streaks, spots or anything that would obscure the viewedimage.6. Check for any stray light that interferes with looking

Pagina 103

8Strap Wrench (T-10913)Apex Torx Bit: Tx 05Hex Key: .035”Pin Vises “C” Size (L.S. Starrett Co. 166c) for Hex KeysTweezers (Grobert Peer #7 St

Pagina 104

1Compact Ophthalmic Set Components:13010 2.5 Volt Pocketscope Ophthalmoscope16201 Pocketscope Retinoscope42010 & 42011 Transilluminator740

Pagina 105

2C1. Retinoscope Body -- Unscrew retinoscope body from handle.Hold knurled ring with thumb of one hand.Unscrew retinoscope with other hand. (It is no

Pagina 106

3Position tape over outside of patient side cover over area that window will seat in.Press the tape down securely. Remove center portionwindow seal

Pagina 107

4D1. Lamp-- Unscrew pocketscope ophthalmoscope from handle.Use tweezers to pull lamp from bottom of ophthalmoscopeoptical tube base.D2. Bumper-- Use

Pagina 108

5D6. Aperture Sub Assy-- Blow off second skid plate.(cont’d) Position it with the ‘ridge’ side downPlace it on top of the dialLift the assembly from t

Pagina 109

6D12. Lens Disc-- Inspect lens disc for dirt/scratches. Remove dust with blowoff gun.Apply Moly33 greaseM11041 sparingly, in thinuniform coat to hub

Pagina 110

71. Operate all slide switches, rotating knobs and discs to determine whether or notdetents are operating correctly and to determine if the part opera

Pagina 111

81. Remove lamp collar--remove lamp from base. Remove the lamp collar from the baseof the Transilluminator. Use tweezers.2. Replace lamp collar-positi

Pagina 112

Traditional Instruments Service Manual PN 117081-1Drawings and/or illustrations and/or part numbers contained in this document are for referencepurpos

Pagina 113

M 11720 Reassembly ContinuedM12 Lens Disc Assy. -- Clean disc. Place disc on hub.(Number side down, engaged with detent)M13 Cover -- Place cover ont

Pagina 114

9Apex torx bit TX-05Hex keys: .035"Pin vises “C” size ( L.S.Starrett co.166c ) for hex keysTweezers ( grobet peer #7 stainless steel )Jewelers

Pagina 115

1767 Wall TransformerService Manual3.5 Volt Wall Transformer Disassembly:Models 76710, 76712, 76714, 76716, 76720, 76722, 76724, 76726, 76730Refer to

Pagina 116 - 60500 1-2

2A. Power Cord ReplacementAbstract:IEC connector accepts new plug. No disassembly. Unplug power cord from bottom ofWall Transformer unit. Refer t

Pagina 117

3Figure 1 - 767 Wiring PC Board Revision AFigure 1 - 767 Wiring PC Board - 767207 Rev. B

Pagina 118

4NOTE:Lay the wires of the right handle assembly underneath the right heatsink. Inspect the wires as they exit from the molex connector to seeif one

Pagina 119

53. Remove Circuit BoardLay the 767 on its back. Remove six 6-32 x 3/4” phillipshead screws and attached plastic washers. Lift board up.(The mountin

Pagina 120

6D. Fuse ReplacementAbstract:Fuses are easily changed by removing the MH-Front and locating the blown fuse on thecircuit board. Inspect for other dam

Pagina 121 - COMPATIBLE NOT COMPATIBLE

7Dirt and or other foreign matter, might beome lodged in the emitter/detector ‘windows’ in the cradle. This could prevent the handle frombecoming ac

Pagina 122

8Figure 2, Cradle Removal

Pagina 123

9F. Illuminated Switch ReplacementAbstract:The MH-Front is removed and the old switch is unlatched and pulled out. The four wiresare disconnected one

Pagina 124

N 11730 3.5 Volt Coaxial Autostep Sealed Ophthalmoscope (With Multiplier) Disassembly:N1 Bumper -- Pull corner, peel back.N2 Back Cover -- Loosen ret

Pagina 125

103. Replace Clock Module in Face-PlatePosition buttons on module with holes in face-plate. Alignclock module with opening in face-plate. Attach th

Pagina 126

113. Plug in the HandleNote the colors of the wires on the plug of the new handleassembly. Match the colors of the wires with the color codingprint

Pagina 127

12Third Handle Checks1. Use main unit as a power supply for these checks. Provide nominal voltage (7.0 V RMS)to third handle assembly.2. Handle outp

Pagina 128

13AppendixInspections/Tests for Repaired 767 TransformersMain Unit and Third HandleIMPORTANT: When any primary component is replaced, perform the lea

Pagina 129

14Third Handle Checks1. Use main unit as a power supply for these checks. Provide nominal voltage (7.0 V RMS)to third handle assembly.2. Handle outpu

Pagina 130

15AppendixTools/Materials/SuppliesFlat Blade Screwdriver (1/4”) with 6” Shank(Stanley 65-322)#1 Phillips Screwdriver#2 Phillips Screwdriver with 6” Sh

Pagina 134

N 11730 Reassembly ContinuedN11 Spring Detent -- Place plastic spring detent into position below bumper cut-outof doctor-side cover, smooth side up, d

Pagina 135

O12810, 12813, 12814, 12830 2.5 Volt Pocket Scope Ophthalmoscope Disassembly:SPECIAL NOTE: The12810, 12813, 12814, and 12830 Ophthalmoscopes are simi

Pagina 136

O 12810, 12813, 12814, 12830 2.5 Volt Pocket Scope Ophthalmoscope Disassembly Cont'dO7 Cover Assembly -- Place lever into position in the ( - )

Pagina 137

APPENDIX 1TOOLS/MATERIALS/SUPPLIESOPHTHALMOSCOPE: 11110, 11400, 11411, 11470, 11475, 11500, 11511, 11600, 11605,11610, 11620, 11630, 11650, 11660, 117

Pagina 138

APPENDIX 2Inspections for Repaired Ophthalmoscopes:1. Operate all slide switches.2. Rotate knobs and discs to determine whether or not detents are ope

Pagina 139

APPENDIX 3Checklist for:Final Inspection of OphthalmoscopesOphthalmoscope Checklist: Ref#: Model:1. [] Bumper in place A02387 (V1)* all2. [] Retaining

Pagina 140

APPENDIX 4Troubleshooting OphthalmoscopesModel #'s: 11400,11411, 11610, 11620,11630, 11650,11660ProblemDirty OpticsCause Corrective ActionNormal

Pagina 141

APPENDIX 4 continued...Troubleshooting OphthalmoscopesMODEL #:11400,11411,11610 11620,11630,11650,11660ProblemNumbers not readable onens assembly.No

Pagina 142

TABLE OF CONTENTSSection 1 ... OphthalmoscopeSection 2 ...

Pagina 143

APPENDIX 4 continued. . . .Troubleshooting OphthalmoscopesMODEL #: 12810 (PocketScope)ProblemMirror unglued, broken ,scratchedCauseMirror adheres to R

Pagina 144

Section 2STRABISMOSCOPEMODEL # PAGE12400 1-6

Pagina 145

STRABISMOSCOPE12401 3.5V Strabismoscope (OCCLUDER) Disassembly:1. Lamp — Pull lamp from base. (Use tweezers)2. Beam Splitter —Loosen retaining ring al

Pagina 146 - LIGHT HANDLES

12401 Reassembly Continued...5. Beam Splitter Reassembly cont'd Determine which side of beam splitter is coated byplacing a pencil point onto

Pagina 147

12401 Reassembly Continued...5. Beam Splitter Reassembly ContinuedFIG. 5Assemble both cover halves together by first engagingthe latch at the top

Pagina 148

STRABISMOSCOPETOOLS/MATERIALS/SUPPLIESBayonet Pliers Welch Allyn T-10802Only one of the next two T-tools is required:Rubber Jaw Ring Pliers Welch Ally

Pagina 149

Inspections For Repaired Strabismoscopes:1. Check locking ring for tightness.2. Check to see that instrument fits on handle before repairing. Base r

Pagina 150

STRABISMOSCOPE TROUBLESHOOTINGMODEL #: 12401Problem Cause Corrective Action1 Broken beam splitter (Mirror)1 Dropped1 Replace beam splitter.IIIIIIIApar

Pagina 151

Section 3RETINOSCOPESMODEL # SUB-SECTION PAGE18201 A 1-217201 B 218100 C 3-417710 D 4-517610 F 618000 F 618010 E 617600 G 7-817700 H 8-9

Pagina 152

RETINOSCOPESRetinoscope Disassembly:A. 18201 3.5 Volt Retinoscope Disassembly:A1. Lamp — Pull lamp from bottom. (Use tweezers).A2. Bumper — Pull cor

Pagina 153

Section 1OPHTHALMOSCOPESMODEL # SUB-SECTION PAGE11110 A 1-2,11400, 11411 B 311470, 11475, C 411500, 11511 D 5-611600 E 711605 F 811610 G 911620 H 1011

Pagina 154

18201 Retinoscope Reassembly ContinuedA9. Doctor-side Cover __ Install doctor-side window and window retainer. Join doctorside and patient side assem

Pagina 155

C. 18100 3.5. Volt Halogen Streak Retinoscope Disassembly:C1. Lamp Carrier — Pull lamp carrier from bottom. (Use needle-nose pliers).Inspect small sp

Pagina 156

18100 Retinoscope Reassembly ContinuedC12. Lamp Carrier — Position lamp carrier into control sleeve with very lightpressure and slight rotation to ali

Pagina 157 - Section 17

17710 Retinoscope Reassembly ContinuedD8. Head Assembly — Screw sleeve assembly into collar of head assembly. (Usestrap wrench to tighten). Align as p

Pagina 158

E. 18010 3.5. Volt Halogen Spot Retinoscope Disassembly:E1. Lamp — Pull 3.5 v. Lamp from base. (Use needle nose pliers).E2. Bumper — Pull corner, pe

Pagina 159

G. 17600 Spot Retinoscope Disassembly:G1. Lamp Carrier — Unscrew light carrier from body subassembly.G2. Lamp — Remove diaphragm assembly. Unscrew la

Pagina 160

17600 Reassembly ContinuedG14. Lamp Carrier — Screw light carrier to body subassembly.Test.End 17600 Reassembly—————————————————————————————————H. 17

Pagina 161

17700 Reassembly ContinuedH11. Bumper — Position bumper over top of cover and attach cover to bodysub-assembly with 3 screws (2 short screws at top).

Pagina 162

APPENDIXRETINOSCOPES TOOLS/MATERIALS/SUPPLIESNeedle Nose Pliers.028 Hex KeyTorx .05 ScrewdriverStrap Wrench Welch Allyn T-10913Slide Switch Holding

Pagina 163

Inspections For Repaired Retinoscopes1. Check streak focus and out of focus uniformity.2. Check polarizer switch for positive detent and smooth opera

Pagina 164

OPHTHALMOSCOPESSpecial Instructions Ophthalmoscope Service:Clean all returned ophthalmoscopes by wiping with a solution of 10% Clorox / 90% water.Do n

Pagina 165

TROUBLESHOOTING RETINOSCOPESInstrument: Retinoscopes (Streak)Model #: 18100Problem Cause Corrective ActionControl sleeve operation difficult Light car

Pagina 166

TROUBLESHOOTING RETINOSCOPESInstrument: RetinoscopesModel #: 18200 streak and 18300 spotProblemFilm on window (difficult to seeimage)Not focusing prop

Pagina 167

Section 4OTOSCOPESMODEL # SUB-SECTION PAGE20000 A 123520 A 124020 B 223510 B 223570 B 224033 B 224031 B 225020 B 221110 C 2-321130 D 320100 E 4-620150

Pagina 168

OTOSCOPESSPECIAL INSTRUCTIONS: Refer to corresponding diagrams in Welch Allyn Repair Parts Catalog PN 900299. Clean allreturned scopes by wiping with

Pagina 169

B. 24020 (24033 without Speculum) 2.5V Ototscope Disassembly:24033 (24031 without Speculum) 2.5V Otoscope Disassembly:25020 (25022 without Speculum

Pagina 170

C. 21110 Otoscope Disassembly ContinuedC9. Otoscope — Screw illuminator assembly onto battery handle.IlluminatorAssemblyC10. Otoscope Body — Screw ot

Pagina 171

E. 20100 (20101without Speculum) 2.5V Otoscope Disassembly:20150 2.5V OotoscopeDisassembly:(Above otoscopes are WIRED and not fiber-optic.)E1. Lamp —

Pagina 172

20100, 20150 Otoscope Reassembly ContinuedFIG. 1 FIG. 2 FIG. 3E6. Bayonet Assembly __ Insert wire through the bayonet assy. Fill with medic

Pagina 173

20100, 20150 Otoscope Reassembly ContinuedE8. Lens — Install lens in window assembly and secure with lens retainer.Screw and tighten lens retainer wit

Pagina 174

G. 20202 (20203 without Speculum) 3.5V Otoscope Disassembly:20262 (20263 without Speculum) 3.5V Otoscope Disassembly:G1. Lamp — Pull lamp out of b

Pagina 175

A. 11110 2.5 V Halogen Lite Set™ Ophthalmoscope Cont'd:A5 Ophthalmoscope -- Remove: Horseshoe (PN111026)Body Assembly Numeral IlluminatorApertu

Pagina 176

20202, 20262 Reassembly ContinuedG11. Double Viewer — Install new ‘o’ ring on body & light pipe assembly. AttachAssembly double viewer subassemb

Pagina 177 - Installation of:

**** 21600 Reassembly ****H6. Contact Assembly — See FIG. 5. Position the wire downwards relative to thereplacement contact assembly. Insert the wi

Pagina 178

21600 Reassembly ContinuedFIG. 7H9. Lock Nut — Place lock nut over body assembly stem. Insert speculum holderinto stem of body assembly in ‘up’ positi

Pagina 179

I. 21700 (21701) 3.5V Otoscope:I1. Lamp — Pull lamp out of base.I2. Lens Holder — Unscrew lens holder screw. Remove lens holder spring washer,Assembly

Pagina 180

APPENDIXTOOLS/MATERIALS/SUPPLIESOTOSCOPEBayonet Pliers (Welch Allyn T-10802)Lens Window Wrench (Welch Allyn T-847)Strap Wrench (Welch Allyn T-10913)Fl

Pagina 181

APPENDIXInspections for Repaired Otoscopes:1. Check fit of disposable and re-useable speculums.2. Check to see that instrument fits on handle before r

Pagina 182

TROUBLESHOOTING OTOSCOPESInstrument: OtoscopeModel #: 20000, 24020, 25020, (21110, 21130 Pocket)ProblemLow or no lightShadowsNot air tightBody apartS

Pagina 183

-TROUBLESHOOTING OTOSCOPESInstrument: Otoscope (Pneumatic)Model #: 20100 (2.5 Volt), 20200 (3.5 Volt)Problem CauseCorrective ActionNo light Blown lamp

Pagina 184

TROUBLESHOOTING OTOSCOPESInstrument: OtoscopeModel #: 21600, 217OOProblem Cause Corrective ActionLow or no light Blown lamp Replace lampLight flicker

Pagina 185

Otoscope Pressure Testing form A02800Otoscope Pneumatic Test (Using Fixture T7503)- Connect specula tip to otoscope.- Connect surgical tubing to ins

Pagina 186

B 11400 and 11411 2.5 Volt Ophthalmoscope Disassembly:B1 Bumper -- Pull corner, peel back.B2 Back Cover -- Loosen and remove retaining ring.(Doctor

Pagina 187

Otoscope Pressure Testing from A00181Equipment:1. Low volume air pump (Aquarium style or equal)2. Bleeder valve3. Flow control4. ”U” tube 12" min

Pagina 189

Section 5ILLUMINATED BIVALVE NASAL SPECULUMMODEL # PAGE26000 1-3

Pagina 190

Illuminated BiValve Nasal SpeculumSpecial Instructions:Refer to corresponding diagrams in Welch Allyn Repair Parts Catalog 900299. Clean allreturne

Pagina 191

26000 Reassembly Continued...6. Bayonet Assembly — Insert wire through the bayonet assembly. Screw the bayonetassembly into the base while carefull

Pagina 192

APPENDIXTOOLS/MATERIALS/SUPPLIESNasal Specs 26000Bayonet Pliers (Welch Allyn T-10802)Needle Point AwlFlat Blade Screwdriver ie. Xcelite # P-181Needl

Pagina 193

Section 6LARYNX ILLUMINATORNASOPHARYNX ILLUMINATORMODEL # PAGE27000 1-227050 1-227200 1-227250 1-2

Pagina 194

IlluminatorSpecial Instructions:Refer to corresponding diagrams in Welch Allyn Repair Parts Catalog 900299. Clean allreturned instruments by wiping wi

Pagina 195

TOOLSTweezers ( Grobet Peer #7 )TROUBLESHOOTING ILLUMINATORSInstrument: Larynx IlluminatorModel #:27000,27200Nasopharynx27050,27250ProblemBroken, bent

Pagina 196

Section 7TONGUE DEPRESSORMODEL # SUB-SECTION PAGE28000 A 1-228100 B 2-3

Pagina 197

C 11470 and 11475 2.5 Volt Ophthalmoscope Disassembly:C1 Bumper -- Pull corner, peel back.C2 Back Cover -- Loosen and remove retaining ring.(Doctor

Pagina 198

TONGUE DEPRESSORSpecial Instructions: Refer to corresponding diagrams in Welch Allyn Repair Parts Catalog 900299. Clean allreturned instruments by w

Pagina 199

28000 Reassembly ContinuedA6. Pipe Assembly — Position the pipe assembly up to the back of the chromebarrel and secure it by tightening the retaining

Pagina 200

**** 28100 Reassembly ****B5. Contact Assembly __ Assemble rubber tube onto wire of contact assembly andinsert contact assembly into retaining screw.I

Pagina 201 - Service Manual

Tongue DepressorsAPPENDIXTOOLS/MATERIALS/SUPPLIESBayonet Pliers T-10802Ring Nut Wrench For 28000. (Single’TEE’ “Wiring Tool” Drawing #23500877) T-1128

Pagina 202 - CAUTION > > >

Section 8PROCTO-SIGMOIDOSCOPES:FIBER OPTIC ANOSCOPESPROCTO-SIGMOIDOSCOPESFIBER OPTIC LIGHT HEADSTANDARD ANOSCOPESROTATING ANOSCOPESMODEL # SUB-SECTION

Pagina 203

Fiber Optic AnoscopesProcto-SigmoidoscopesFiber Optic Light HeadStandard AnoscopesRotating AnoscopeSpecial Instructions: Refer to diagrams in Welch Al

Pagina 204

B. 36019 Fiber Optic Light Head Disassembly:B1. Window Assembly — Remove spirol pin. Collect 2 plungers and 2 plungersprings as you slowly remove wind

Pagina 205

D. 38108, 38114, 38119, 38122 Standard Anoscopes Disassembly:D1. Obturator Assembly — Pull obturator out of funnel assembly reassembly.D2. Obturator A

Pagina 206

38800 Rotating Anoscope Reassembly ContinuedE8. Contact Assembly __ Put contact assembly wire through contact assemblyspring (PN 383065) and position

Pagina 207

Procto Sigmoidoscopes/AnoscopesAPPENDIXTOOLS/MATERIALS/SUPPLIESWindow Screw Wrench Large (Undercut) ( Welch Allyn T-5320 )Window Screw Wrench Small (U

Pagina 208 - Figure 2, Cradle Removal

D 11500 and 11511 2.5 Volt Ophthalmoscope Disassembly:D1 Rubber Hood -- Pull corner, lift up.D2 Light Carrier -- Pry/pull out light carrier assembly

Pagina 209

TROUBLESHOOTING PROCTO SIGMOIDOSCOPES/ANOSCOPEInstrument: Rotating Anoscope With 38300 Light CarrierModel #: 38800Problem Cause Corrective ActionRing

Pagina 210

Section 9HEADLIGHTSAND TRANSFORMERMODEL # SUB-SECTION PAGE46003 A 1-246023 A 1-246050 A 1-249003 B 3-449022 B 3-449030 B 3-449034 B 3-449035 B 3-44903

Pagina 211

Headlightsand TransformerThe terms POWER SUPPLY, TRANSFORMER, and AC ADAPTOR, are used interchangeably.Special Instructions:Clean returned headlights

Pagina 212

A. Direct Focusing Headlight NOS: 46003 (120Volt), 46023 (220Volt), 46050 (250Volt)Disassembly Continued:A4. Male Switch — Un-solder two wires from b

Pagina 213 - Appendix

Halogen Headlight: 49003 (110-130V) & 790030-501*Power Supply49022 (220-250V) & 790030-502 Power Supply49030 & 470305 IEC AC Adaptor*4903

Pagina 214

B. Halogen Headlight/IEC AC adaptors: 49003, 49034, 49035, 49036Disassembly: (Transformers for the 490 series are serviceable items, refer to Section

Pagina 215

790030-501 (110-130Volt) & 790030-502 (220-250Volt) Power Supplies for 49003 & 49022Headlight RespectivelyC DisassemblyC1. Rocker Switch — Di

Pagina 216

790030-501 (110-130Volt) & 790030-502 (220-250Volt) Power Supplies for 49003 & 49022Headlight respectivelyC Disassembly ContinuedC3. Transfor

Pagina 217

470305, 470225, 470245, 470285 IEC AC ADAPTORS for 49030, 49034, 49035, 49036Headlights respectivelyD DisassemblyD1. Rocker Switch — Disconnect IEC AC

Pagina 218

470305, 470225, 470245, 470285 IEC AC Adaptors for 49030, 49034, 49035, 49036Headlights respectivelyD Disassembly ContinuedD5. Female Switch — Unsc